Application: It is used to mix into dialysis solution in the treatment of acute/chronic renal failure during hemodialysis and drug poisoning.
Features:
1.All substances used conform to requirements of relevant standard and are produced in strict production environment and undergone rigorous testing.
2.The electrolytes of potassium, calcium and magnesium ion and glucose in formula can be adjusted in the reasonable range according to requirements from varialbe machine such as Fresenius,Gambro,Baxter, B.Braun, Nikkiso, Nipro etc.
3. It can be produced according to the formula provided by customers.
Cautions:
1.Dissolve and dilute dry Concentrate by RO water (meet AAMI stand)
2.Acid concentrate and Bicarbonate concentrate can’t be mixed directly or used separately. Concentrate solutions and water are proportioned by dialysis machine and mixed on line acc. to predetermined dilution ratio
3.The dialysis solution can’t be used for peritoneal dialysis and injection
4.Do not use the dissolved concentrate solution in case of color change and precipitationFormula for examples
Features:
1.All substances used conform to requirements of relevant standard and are produced in strict production environment and undergone rigorous testing.
2.The electrolytes of potassium, calcium and magnesium ion and glucose in formula can be adjusted in the reasonable range according to requirements from varialbe machine such as Fresenius,Gambro,Baxter, B.Braun, Nikkiso, Nipro etc.
3. It can be produced according to the formula provided by customers.
Cautions:
1.Dissolve and dilute dry Concentrate by RO water (meet AAMI stand)
2.Acid concentrate and Bicarbonate concentrate can’t be mixed directly or used separately. Concentrate solutions and water are proportioned by dialysis machine and mixed on line acc. to predetermined dilution ratio
3.The dialysis solution can’t be used for peritoneal dialysis and injection
4.Do not use the dissolved concentrate solution in case of color change and precipitationFormula for examples
Code | mEq/L | g/L | Applicable Machine | Dilution Ratio | PH | ||||||
Na+ | K+ | Ca++ | Mg++ | Cl- | HCO3- | CH3COO- | C6H12O6 | ||||
TMSDCF01 | 138.0 | 2.5 | 1.5 | 0.50 | 109.5 | 32.0 | 3.0 | Fresenius | 1:1.225:32.775 | 7.0~7.8 | |
TMSDCB01 | 138.0 | 2.0 | 1.50 | 0.50 | 108.0 | 29.0 | 8.0 | 1.0 | B.braun, Nikkiso | 1:1.26:32.74 | 7.0~7.6 |
TMSDCG01 | 140.0 | 2.0 | 1.75 | 0.50 | 108.0 | 34.0 | 5.0 | Gambro | 1:1.83:34 | 7.0~7.6 | |
TMSDCN01 | 136.0 | 2.5 | 1.50 | 0.75 | 109.0 | 30.0 | 3.0 | 1.0 | Nipro | 1:1.26:32.74 | 7.0~7.6 |
TMSDCB01 | 138.0 | 2.5 | 1.75 | 0.50 | 108.0 | 35.0 | 4.0 | Baxter | 1:1.83:34 | 7.0~7.6 | |
TMSDCT01 | 135.0 | 2.5 | 1.75 | 0.50 | 106.0 | 30.0 | 8.0 | 1.0 | Toray | 1:1.26:32.74 | 7.0~7.6 |
Remarks:
The formula can be adjusted according to the requirements for clinical use.
Packaging:
The formula can be adjusted according to the requirements for clinical use.
Packaging:
1Acid concentrate powder : one dialysis sessions volumn per poly bag and then several polybag be packed into one carton or paper bag
2 Bicarbonate concentrate powder one dialysis sessions volumn per poly bag and then several polybag be packed into one carton or paper bag.
2 Bicarbonate concentrate powder one dialysis sessions volumn per poly bag and then several polybag be packed into one carton or paper bag.
3.Expiry date: 1 year.
没有评论: